Specifications, General, Dimensions – GE P&W Ultrapure and Drinking Water TOC Analyzers - Sievers Nitric Oxide Analyzer (NOA 280i) User Manual
Page 4

Specifications
Specifications
Liquid Samples
Gas Samples
Sensitivity
1 picomole (1 nM for a 1 mL injection)
0.5 ppb
Range
Nanomolar to millimolar
0.5 ppb-500 ppm
Repeatability
+/- 5%
+/- 5%
Sample
Volume: 0.001–1 mL
Flow rates: 10-200 mL/min
General
Response Time
Electronics: 67 milliseconds to 90% full scale. Lagtime: 1 second
Outputs
(1) analog 0-1V, 0-10V (1) digital RS-232 (9600-38400 Baud), (1) printer parallel port
Data Sampling Rate
Operator selectable — 32, 16, 8, 4, 2,or 1 samples per second; or 12, 4, 2, 1, 0.2, or
0.1 samples per minute
Power Requirements
U.S. — 110/115V, 60 Hz (6A)
Japan — 90/100V, 50/60 Hz (7A)
Europe, Asia and South America — 220/240V 50 Hz (3A)
Dimensions
NOA 280i
41 H x 16 W x 51 L cm (16 H x 6.2 W x 20 L in)
Weight: 16 kg (35 lb)
Vacuum Pump
37 H x 19 W x 48 L cm (14.5 H x 7.5 W x 19 L in)
Weight: 21.5 kg (47 lb)
* Trademark of General Electric Company; may be registered in one or more countries.
1
US Patent Numbers: 5,795,787 and 6,010,459. Under license from Aperon Biosystems, Inc.
Caution: For Research Use Only. Not for Use in Diagnostic Procedures.
©2008, General Electric Company. All rights reserved.
300 00089 Rev B, Jun-08
MC08-131
The Americas
GE Analytical Instruments
6060 Spine Road
Boulder, CO 80301-3687 USA
T +1 800 255 6964
T +1 303 444 2009
F +1 303 527 1797
Europe/Middle East/Africa
GE Analytical Instruments
Unit 3, Mercury Way
Urmston, Manchester
UK M41 7LY
T +44 (0) 161 864 6800
F +44 (0) 161 864 6829
www.geinstruments.com
Asia Pacific
GE Analytical Instruments
5 United Plaza
1468 Nanjing Rd (W)
Shanghai 200040
China
T +86 021 3222 4555