VEGA PLICSCOM User Manual
Page 21

21
5 Setup - Sensors up to software 3.99
PLICSCOM •
36433-EN-130228
The following values can be displayed:
•
Height
•
Pressure (only with pressure transmitters)
•
Distance (only with radar, guided microwave, ultrasonics)
•
Current
•
Scaled
•
Percent
•
Lin. percent
•
Temperature (only with pressure transmitters).
The selection "scaled" opens the menu items "Display unit" and
"Scaling". In "Display unit" there are the following options:
•
Height
•
Ground
•
Flow
•
Volume
•
Without unit
Depending on selection, the different units are in turn available.
In the menu item "Scaling", the requested numerical value with deci-
mal point is entered for 0 % and 100 % of the measured value.
There is the following relationship between the indication value in the
menu "Display" and the adjustment unit in the menu "Device settings":
•
With radar, guided microwave and ultrasonics, the displayed value
"Distance" means: presentation of the measured value in the
selected adjustment unit, e.g. m(d)
•
With pressure, the displayed value "Pressure" or "Height" means:
presentation of the measured value in the selected adjustment
unit, e.g. bar or m.
Displayed value
Scaled
▼
Display unit
Volume
▼
l
▼
Scaling
0 % = 0.0 l
100 % = 100.0 l
In the menu item "Current output" you determine the behaviour of the
current output during operation and in case of failure. The following
options are available:
Service/Current output