12 supplement, 1 technical data – VEGA VEGAMET 391 With SIL qualification User Manual
Page 46
46
12 Supplement
VEGAMET 391 • With SIL qualification
38704-EN-130228
12 Supplement
12.1 Technical data
General data
Version
Instrument for mounting into front panel, switching cabi-
net or housing
Weight
620 g (1.367 lbs)
Housing materials
Valox 357 XU
Connection terminals
Ʋ Type of terminal
Pluggable spring-loaded terminal with coding
Ʋ Max. wire cross-section
2.5 mm² (AWG 14)
Voltage supply
Operating voltage
20 … 253 V AC, 50/60 Hz, 20 … 253 V DC
Max. power consumption
7 VA; 3 W
Sensor input
Number of sensors
1 x 4 … 20 mA
Input type
Ʋ Active input
Sensor supply through VEGAMET 391 mit SIL-Qualifi-
kation
Measured value transmission
Ʋ 4 … 20 mA
analogue for 4 … 20 mA sensors
Deviation
Ʋ Accuracy
±20 µA (0.1 % of 20 mA)
Terminal voltage
19 … 14.5 V at 4 … 20 mA
Current limitation
approx. 26 mA
Detection line break
≤ 3.6 mA
Detection shortcircuit
≥ 21 mA
Adjustment range 4 … 20 mA sensor
Ʋ Empty adjustment
2.4 … 21.6 mA
Ʋ Full adjustment
2.4 … 21.6 mA
Ʋ min. adjustment delta
16 µA
Connection cable to the sensor
two-wire screened standard cable
Digital input
Quantity
2 x digital input
Input type
Passive
Switching threshold
Ʋ Low
-3 …5 V DC
Ʋ High
11 …30 V DC
Max. input voltage
30 V DC
Max. input current
30 mA