3 technical data, transducer – VEGA VEGASON 71 - D User Manual
Page 7

VEGASON 71 - D
7
Level module
2 relay outputs
floating spdts each
and 2 status indications (LEDs)
- contact material
AgCdO and Au plated
- min.
turn-on voltage
10 mV
switching current
10
µ
A
- max.
turn-on voltage
250 V AC, 60 V DC
switching current
2 A AC, 1 A DC
- max.
breaking capacity
125 VA, 60 W
Current module
2 current outputs
- range
0/4 … 20 mA
- resolution
0,05 % of range
- load
max. 500 Ohm
- load dependent error
at 0 … 500 Ohm, < 0,2 % related to the range
or module
as described above
however with 2 floating outputs each
Fault signal
1 fail safe relay
1 floating spdt
contact data
as described under relay outputs
2 failure-LEDs
2.3
Technical data, transducer
Transducer SW 71
Material of the transducer housing
PVDF
Material of the fixing tube
PVDF, thread G 1 A
Connection cable to central electronics
standard coaxial cable type RG 58
Cable length
5 m, as option 300 m
Diameter of cable
approx. 5 mm
Temperature sensor
integrated in the transducer
Ambient temperature
-20
°
C … +80
°
C
Storage and transport pressure
-20
°
C … +80
°
C
Protection
IP 68
Permissible ambient temperature
1 bar
Weight
approx. 0,8 kg
Transducer SW 71 R Ex
Ex-approval
PTB-no. Ex-94.C.4066
Flame proofing
Casting "m"
Marking
EEx m II T6 or T5 or T4
Material of the transducer housing
PVDF
Material of the fixing tube
PVDF, thread G 1 A
Connection cable to central electronics
standard coaxial cable type RG 58
Cable length
5 m, as option up to 300 m
Diameter of cable
approx. 5 mm
Temperature sensor
integrated in the transducer
Temperature class
T6 = 60
°
C, T5 = 75
°
C, T4 = 80
°
C
Storage and transport temperature
-20
°
C … +80
°
C
Protection
IP 68
Permissible ambient pressure
1 bar
Weight
approx. 1,0 kg
2 Technical Information