6 Flow measurement

6.2

Linearity protocol

Linearization curve

Index

Level percent

Flow percent

marker no.

(level)

01

02

03

04

05

06

07

08

09

10

11

12

13

14

15

16

17

18

19

20

21

22

23

24

25

26

27

28

29

30

31

32

Date

Name

6.1

Linearization (continuation)

Programming example

When programming the level and flow percent values,
observe the following:
- The starting point of the characteristics (index marker

00) is generated by the instrument itself and
coordinated with the values H = 0.0, L = 0.0.

- The end (last index marker) can be made individually

within the 32 value pairs.

- The last index marker must only consist of

100.0

level

percent and

100.0

flow percent.

Programming acc. to linearity protocol

(see e.g. page 21)

Index marker 01 (level percent)
• Enquire index marker

4.H.01

in the MODEFIELD

• Program

30

% in the PARAMETER FIELD acc. to the

linearization curve

• Then store with STO

Index marker 01 (flow percent)
• Enquire index marker

4.L.01

in the MODEFIELD

• Program

10

% in the PARAMETER FIELD acc. to the

linearization curve

• Then store with STO

etc. acc. to linearity protocol. After programming of all data
- reset to mode range 0
- change display indication to flow proportional indication,

i.e. mode

0 - 13

0 - 16

- here in the example activate the programmed

linearization curve, i.e. mode

0 - 18

=

4

.