7 Output results / 8 Supplementary programmings
8
Supplementary programmings
8.1
Failure processing
In mode
0 - 93
the reaction of the relay and current outputs
in case of failure can be defined.
Programming possibilites
Mode
Current outputs
Relay outputs
0 - 93
LED
= 1
actual currents are
no change
stored
= 2
current - 0 mA
= 3
current corresp. to 0 %
operated relay de-energize
= 4
current corresp. to 100%
Programming example
Change-over to failure processing
e.g.
0 - 93
=
4
• Enquire mode
0 - 93
in the MODEFIELD
• Program figure
4
in the PARAMETER FIELD
• Then store with STO
Note
For adjustment of the failure processing observe
additionally the schedule of error codes in the supplement.
7.8
Current module
The pulse-echo measuring system can be equipped with
another module. The module is provided with two current
outputs.
Current output 1 is automatically coordinated to the flow
proportional output result and current output 2 to the level
proportional output result.
The course of characteristics of the current output is fixed
by an initial and final point.
The factory setting defines a course of 4 … 20 mA
corresponding to 0 … 100 %.
Within the whole range of 0.00 … 20.00 mA, each current
value can be programmed for a raising or falling
characteristics.
The current-
∆
between initial and final value must be min.
1 mA.
Programming example
Factory setting
Programming example
100
80
60
40
20
16
0
0
% mA
mA
20
4
End of characteristics 100 % ^ 16 mA
•
Enquire mode
0 - 65
in the MODEFIELD
•
Program
16.00
mA in the PARAMETER FIELD
•
Then store with STO
Begin of characteristics 0 % ^ 0 mA
•
Enquire mode
0 - 66
in the MODEFIELD
•
Program
00.00
mA in the PARAMETER FIELD
•
Then store with STO
Mode
0 - 66
Begin of
characteristics
Mode
0 - 65
End of
characte-
ristics