VEGAMET 515
Product description
Ambient conditions
Permissible ambient temperature
-20°C … +60°C
Storage and transport temperature
-40°C … +80°C
Electrical connection
Multipoint connector component
acc. to DIN 41 612, series F, 33-pole, 3 rows
d, b, z (partly equipped)
Carrier BGT 596 (Ex)
connection to appropriate module
Housing type 505
connection to screw terminals (max. 2.5 mm
2
)
Electrical protective measures
Protection
- unassembled
IP 00
- mounted into
carrier BGT 596
- front side completely equipped
IP 40
- upper and lower side
IP 20
- wiring side
IP 00
- mounted into
housing type 505 or 506
- terminal side
IP 20
- housing in general
IP 30
Protection class
II (in housing type 505)
Overvoltage category
II
Electrical separating measures
Reliable separation acc. to
VDE 0106, part 1 between
power supply, fail safe relay and level relay
and meas. data inputs
- reference voltage
250 V
- insulation resistance
2.3 kV
Galvanic separation
among relay outputs
- reference voltage
250 V
- insulation resistance
1.4 kV
Potential separation
between DISBUS and outputs
- reference voltage
50 V
- insulation resistance
0.5 kV
Common reference potential on
voltage and current output, correction signal
input
Mechanical data
Series
module unit for carrier BGT 596
or housing type 505
Dimensions
W = 25.4 mm (5 TE), H = 128.4 mm, D = 162 mm
CE conformity
VEGAMET 515 and VEGAMET 515 Ex signal conditioning instruments meet the protective
regulations of EMC (89/336/EWG) and NSR (73/23/EWG). Conformity has been judged
acc. to the following standards:
EMC
Emission
EN 50 081 - 2: 1993
Susceptibility
EN 50 082 - 2: 1995
NSR
EN 61 010 - 1: 1993