VEGA VEGAMET 515 User Manual
Page 7
VEGAMET 515
7
Voltage output
Number
3 outputs
Function
analogue output of the processing results
Range
individually adjustable in the range of 0 … 10 V
Current
max. 1 mA
Resolution
0.5 mV
Linearity error
0.05 % (relating to 10 V)
Temperature error
0.06 %/10 K (relating to 10 V)
Relay outputs
Number
2 switching relays
1 fail safe relay
Contact
floating spdt
Contact material
AgNi, hard gold plated
Turn-on voltage
min. 10 mV DC
max. 250 V AC/DC
Switching current
min. 10 µA
max. 3 A AC, 1 A DC
Breaking capacitance
max. 500 VA, 54 W
Min. switching hysteresis
(Low-/High-Delta)
0.5 %
DISBUS output
Function
for wiring between the signal conditioning instru-
ments and for connection of digital indicating
instruments
Connection cable
2-wire standard cable (screening
recommended)
max. cable length
1000 m
Indicating elements
Clear text indication
LC display
(with VEGAMET 515N not available)
- 4 lines, 6 digits each
- background lighting
Analogue indication
LED row consisting of:
(with VEGAMET 515N not available)
- 11 segments 0 % … 100 %
- shows the actual value of the selected
measurement loop
LEDs in front plate
green: operating voltage on
red: fault signal (LED lights with deenergised
relay)
yellow: relay switching condition (standard
setting: LED lights with energised relay)
Adjustment elements
Front plate
6 keys for configuration and parameter setting
(with VEGAMET 515N not available)
Circuit board
rotary switch for adjustment of the instrument
address on DISBUS
Product description