VEGA VEGASON 54P…56P Profibus PA User Manual
Page 59
VEGASON 54P … 56P
59
Technical data
Characteristics
1)
(typical values under reference conditions, all statements relate to the nominal measuring
range)
Min. span
(between empty and full adjustment)
> 20 mm (recommended > 50 mm)
Ultrasonic frequency (at 20°C)
- VEGASON 54
30 kHz
- VEGASON 55
18 kHz
- VEGASON 56
10 kHz
Intervals
- VEGASON 54
1.0 s
- VEGASON 55
1.5 s
- VEGASON 56
2.0 s
Beam angle at -3 dB emitted power
- VEGASON 54
4°
- VEGASON 55
5°
- VEGASON 56
6°
Influence of the process temperature
negligible, is compensated by a dynamic
temperature detection system integrated in the
transducer. Without temperature compensation,
the ambient temperature error will be 0.18 %/K
Influence of the process pressure
negligible within the approved sensor pressures
Adjustment time
2)
- VEGASON 54
> 2 s (depending on parameter setting)
- VEGASON 55, 56
> 4 s (depending on parameter setting)
Ex technical data
Classification
m (encapsulation of the transducer)
Temperature class (permissible ambient
temperature on the transducer
when used in Ex areas)
- T6
42°C
- T5
58°C
- T4
60°C
- T3
60°C
Ex approved in category or Zone
- VEGASON 56
ATEX
Zone 1 (II 2G)
IEC, CENELEC, PTB
Zone 1
- VEGASON 54 … 56
ATEX
Zone 21/22 (II 2D/3D)
IEC, CENELEC, DMT
Zone 10/11
1)
Similar to DIN 16 086, reference conditions acc. to IEC 770;
temperature 15°C … 35°C; moisture 45 % … 75 %; pressure 860 mbar … 1060 mbar.
2)
The adjustment time is the time required by the sensor to output a correct level (with max. 10 % deviation)
after a sudden level change.