Veris Industries 228x SERIES Datasheet User Manual
Insertion meter, standard impeller, Ideal for pipe sizes 3” to over 40, Fl ow
![background image](/manuals/259862/1/background.png)
fl
ow
VERIS INDUSTRIES
800.354.8556
+1 503.598.4564
www.veris.com
Insertion Meter,
Standard Impeller
FEATURES
●
Six-bladed impeller design with a proprietary, non-magnetic sensing
mechanism...high accuracy and repeatability
●
Forward-swept impeller is less prone to fouling by water-borne debris...reliable
performance with minimal downtime
●
Signal can travel up to 2000 ft (609 m) between the sensor and the display unit
without amplification
●
Supplied with 20 ft (6 m) of 2-conductor 20-AWG UL type PTLC cable (105°C
rated)
●
Sensors of similar type are interchangeable...no need for recalibration after
servicing or replacement...quick install and fast maintenance
DESCRIPTION
Insert-style liquid flow sensors with brass or stainless steel sleeves fit pipe sizes from
3” to 40” (77 to 1016 mm). These sensors can be purchased with a bronze or iron
tee. Sensor output is a frequency that indicates flow rate. Used in conjunction with a
compatible flow monitor or transmitter, these non-magnetic flow sensors provide an
accurate reading of the rate of liquid flow, as well as total accumulated flow in indoor
or protected areas.
APPLICATIONS
●
Measuring liquid flow rates
●
Accounting for total accumulated flow
DIMENSIONAL DRAWINGS
MODEL
MANUF. PART #
DESCRIPTION
U001-0001
220BR0005-1211
Flow, Sensor, Insert, Brass Sleeve, 3” to 40” (77 to 1016 mm) pipe
U001-0002
220SS0005-1211
Flow, Sensor, Insert, SS Sleeve, 3” to 40” (77 to 1016 mm) pipe
U001-0006
228BR2005-1211
Flow, Sensor, Insert, Bronze, 2” Bronze Tee
U001-0007
228CB2005-1211
Flow, Sensor, Insert, Bronze, 2” Iron Tee
U001-0025
228BR2505-1211
Flow, Sensor, Insert, Brass, 2.5” Bronze Tee
U001-0030
228CB2505-1211
Flow, Sensor, Insert, Brass, 2.5” Iron Tee
U001-0072
228BR2004-0211
Flow, Sensor, Insert, Brass, 2” FMCSA, Viton
SPECIFICATIONS
Temperature Rating
105°C (221°F) continuous
Pressure Rating
At 38°C (100°F) Insert: 400 psi; brass tee: 200 psi; iron tee: 175 psi
Recommended Design Flow Range
0.5 to 30 ft/sec (0.15 to 9 m/sec); initial detection below 0.3 ft/sec
Wetted Materials
UHMW-PE bearing, polyamide impeller, tungsten carbide shaft, EPDM O-rings
Accuracy
1% F.S. over recommended design flow range; ±4% of reading within calibration range*
Repeatability
±0.3% of full scale over recommended design flow range*
Linearity
±0.2% of full scale over recommended design flow range*
Output Frequency
3.2 Hz to 200 Hz
Output Pulse Width
5 msec ±25%
* ≥10 upstream and ≥5 downstream straight pipe diameters, uninterrupted flow.
Ideal For Pipe Sizes 3”
To Over 40”
220BR0005-1211
220SS0005-1211
Year
Warranty
1
ORDERING INFORMATION
228CB2005-1211
3”
(76 mm)
1/2 - 14 NPT
Pipe I.D.
1.5”
(38 mm)
1.75”
(45 mm)
7.2”
(183 mm)
5.25”
(133 mm)
(insertion depth)
2” NPT
9”
(229 mm)
4.5”
(114 mm)
220x & 228x SERIES
HQ0001778.B 01131