6. Retrieving Logged Data
Logged data can be displayed on the 1008 display or sent to a PC by using the
USB cable and software included. The PC software also provides the capability of
real time monitoring from a PC.
Recalling and Displaying Logged Data
Retrieving Logged Data via PC
To retrieve logged data the USB cable and PC software (both supplied) must be
used.
Install the PC software on your PC and open it.
Connect the 1008 to the PC using the USB cable.
Run the software and select 1008 when prompted. Click on the connect to PC
button in the software window.
The software can be used to retrieve logged data and for real time monitoring.
COM port setting is:
Baud rate: 19200
Parity: None
Data Bits: 8
Stop Bit: 1
1) Press the
key, “LOG 01” and the logged CO2 and temperature reading
will alternately flash. Press the
to cycle through all of the logged
data.
2) To deactivate recall mode and return to normal operation, press the
key.
7. Technical Specifications
Function
Range
Accuracy
CO2
0 to 5000ppm
50°F~104°F: ±3% of rdg+75ppm
<50°F, >104°F: ±5% of rdg+75ppm
Temperature
-5°F to 140°F
±2°F
-20°C to 60°C
±1°C
Data Logging: Up to 50 samples 1 minute to 60 minutes in 1, 2, 3, 5, 10, 20, 30,
40, 50, 60 minute intervals
Display Type: 2 line with annunciators and backlight
CO2 Alarm: Audible and visual, selectable level (factory default off)
Operating Temperature & Humidity
CO and CO2: 14°F to 122°F (-10°C to 50°C)
All Other Function: -4°F to 140°F (-20°C to 60°C)
Humidity 15 ~ 95% non-condensing
Storage Temperature: -4°F to 140°F (-20°C to 60°C)
Power supply
3 AA battery( 4.5V)
AC/DC adapter 5Vout (Center positive)
Battery Life: 40 hours typical without backlight
Output: Serial output via USB connection