Operation
Setup Menu : Load
3.2.2 Load Setup Submenu
If the Load submenu is selected in setup, then several parameters 
associated with the superconducting magnet load can be viewed and/or 
specified by using the 
6(783
key to cycle through the available items.
3.2.2.1 Stability Setting
The stability setting is specified in percent and controls the 
transient response and stability of the system. The valid range is 
from 0.0 to 100.0%. The default value is 0.0% unless preset by AMI. 
The chart below may be used as a guide to set the stability setting 
for magnets with a persistent switch installed. Magnets with an 
inductance of greater than 3 Henries that have a persistent switch 
installed should operate without problems with a stability setting 
of 0.0%.
3.2.2.2 Coil Constant
The coil constant is a scaling factor which converts the operating 
current to kilogauss (10 kG = 1 Tesla). It is also often referred to as 
the field-to-current ratio. The coil constant is specified in kilogauss/
$ × 6WDELOLW\ 6HWWLQJ
9V
Model 420 Stability Setting vs. Magnet Inductance
0
10
20
30
40
50
60
70
80
90
100
0
1
2
3
4
5
6
7
Inductance (H)
Sta
b
ility
Se
tting (%
)
$ × &RLO &RQVWDQW N*$
9V