Table 3-2. available select power supply options, Table 3-2, Available – American Magnetics 420 Power Supply Programmer User Manual
Page 59: Operation

41
Operation
Setup Menu : Supply
Note
The operating current must be less than 0.1% of I
max
in order to
change the supply selection. If a change is attempted above this
current, the Model 420 will beep and ignore the keypress. Power
supply selection should also preferably be performed with the power
supply off for maximum safety.
The power supply settings define the V-I ranges for a specific
supply. For example, V-I diagrams are presented in Figure 3-3 for
the AMI 12100PS and AMI 4Q05100PS selections. The AMI
12100PS operates as a one-quadrant system without the addition of
an energy absorber. With the addition of an energy absorber, the
AMI 12100PS system can function as a two-quadrant supply. The
AMI 4Q05100PS power supply operates as a four-quadrant power
supply without the addition of an energy absorber.
The addition of an energy absorber to the system does not change
the capabilities of the power supply (or the values entered for the
supply). The addition of an energy absorber does, however, change
the system operating ranges per the example of Figure 3-3.
Table 3-2. Available Select Power Supply options.
Power Supply
Min Output
Voltage (V)
Max Output
Voltage (V)
Min
Output
Current (A)
Max Output
Current (A)
V-V Mode Input
Range (V)
AMI 12100PS
+0.000
+12.000
+0.000
+100.000
+0.000 to +10.000
AMI 12200PS
+0.000
+12.000
+0.000
+200.000
+0.000 to +10.000
AMI 4Q05100PS
−
5.000
+5.000
−
100.000
+100.000
−
10.000 to +10.000
AMI 10100PS
+0.000
+10.000
+0.000
+100.000
+0.000 to +5.000
AMI 10200PS
+0.000
+10.000
+0.000
+200.000
+0.000 to +5.000
HP 6260B
+0.000
+10.000
+0.000
+100.000
+0.000 to +10.000
Kepco BOP 20-5M
a
−
10.000
+10.000
−
5.000
+5.000
−
10.000 to +10.000
Kepco BOP 20-10M
−
10.000
+10.000
−
10.000
+10.000
−
10.000 to +10.000
Xantrex XFR 7.5-140
+0.000
+7.500
+0.000
+140.000
+0.000 to +10.000
Custom...
b
−
20.000
+20.000
−
200.000
+200.000
−
10.000 to +10.000
a. The Kepco supplies are limited to only 1/2 of the output voltage range since the
supplies are only designed to safely dissipate 1/2 of the rated power output.
b. The values shown for the Custom... option are defaults. The user should enter the
appropriate values within the respective submenus. Custom values, once entered,
are saved between power-ups.