Troubleshooting

A

ROM memory error has occurred. Press the

CLEAR

key to acknowledge the error. If the system does not oper-
ate properly, contact your

100S GP

distributor for repair/

replacement procedures.

A power loss has occurred. This message will appear
everytime you power up the system and can be cleared
by pressing the

CLEAR

key. If the E1 message appears at

any time other than at power up, a power interruption has
occurred and your current position may be lost. Reestab-
lish the datum if necessary. If the problem persists, con-
tact your distributor for repair/replacement procedures.

A random access memory (RAM) error has occurred.
Press the

CLEAR

key to acknowledge the error. If the sys-

tem does not operate properly, contact your

100S GP

dis-

tributor for repair/replacement procedures.

A scale miscount error (invalid waveform) has occurred.
The E9 message will appear on the axis that has the error.
Press

CLEAR

to reset the axis(es). Your current position

may be lost. Re-establish the datum if necessary. If the
system does not operate properly, contact your

100S GP

distributor for repair/replacement procedures.

Error Messages (General)

E1

E2

E3

E4

E9

When

100S GP

is powered up, it will run a series of tests to ensure that

the software is working properly. If an error message appears, refer to
the charts below for a diagnosis and solution.

An EEPROM memory error has occurred. Press the

CLEAR

key to acknowledge the error. The datum and

setup information may be lost. These areas should be
checked and reestablished upon power up. If the problem
persists, contact your distributor for repair/replacement
procedures.