Troubleshooting – ACU-RITE 100S User Manual
Page 24

100S General Purpose Reference Manual
21
Troubleshooting
Scale counts,
but not
correctly
Check that the scale cable
is securely connected to
the readout
Loose cable
Check that the scale cable
is connected to the correct
input on the back of the
readout.
Wrong input
Check the scale and read-
ing head for proper instal-
lation (refer to scale
installation instructions).
Scale not installed
properly
Scale won’t
count
Check the scale and read-
ing head for proper instal-
lation (refer to scale instal-
lation instructions). Clean
the scale if necessary.
Scale not installed
properly
Change the count direction
(pg. 9).
Wrong count
direction
Unplug the scale from the
readout and plug it into an
input for another axis.
If the scale still does not
count, the reading head
may need to be replaced.
If the scale counts on the
other axis then the display
may need to be replaced.
See LED Failure
The scale’s reading
head is not working
Symptom
Probable Cause
Recommended
Corrective Action