100S General Purpose Reference Manual

21

Troubleshooting

Scale counts,
but not
correctly

Check that the scale cable
is securely connected to
the readout

Loose cable

Check that the scale cable
is connected to the correct
input on the back of the
readout.

Wrong input

Check the scale and read-
ing head for proper instal-
lation (refer to scale
installation instructions).

Scale not installed
properly

Scale won’t
count

Check the scale and read-
ing head for proper instal-
lation (refer to scale instal-
lation instructions). Clean
the scale if necessary.

Scale not installed
properly

Change the count direction
(pg. 9).

Wrong count
direction

Unplug the scale from the
readout and plug it into an
input for another axis.
If the scale still does not
count, the reading head
may need to be replaced.
If the scale counts on the
other axis then the display
may need to be replaced.
See LED Failure

The scale’s reading
head is not working

Symptom

Probable Cause

Recommended
Corrective Action