. The aperture iris diaphragm for use in ordinary fluorescence observation is not provided. If it is required to use the aperture

iris diaphragm, replace the slit iris diaphragm unit of this system with the aperture iris diaphragm unit of the IX2-RFAL.

. When the IX2-RFAC, IX2-RFACA and a standard mirror unit are used in combination, it is required to readjust the centering

of the slit iris diaphragm unit because, otherwise, the contrast of the total reflection fluorescent light would fluctuate due
to variation in the illumination light projection positions between turret positions.

. The IX2-SFR right-hand knob stage is not recommended because the stage knobs will degrade the operability of the slit

iris diaphragm unit considerably. Instead, it is recommended to use the IX-SVL-2.

. When the IX2-RSPC right-side port adapter is used, some positions of the excitation filter slider would be unusable due

to interference with the camera. The following table shows the relationship between the camera size and unusable filter
positions.

13

14

15

16

}The camera interference can be reduced by using the U-FWR filter wheel or a commercially available motorized filter

wheel.

Height from the center of
C-mount

Depth from the center of
C-mount

Interference condition

Less than 27 mm

None

27 mm to 45 mm exclusive

Less than 34 mm

None

34 mm to 70 mm exclusive

Deepest (vacant) position is unusable.

70 mm or more

The two deepest positions are unusable.

45 mm or more

Less than 20 mm

None

20 mm to 56 mm exclusive

Deepest (vacant) position is unusable.

56 mm or more

The two deepest positions are unusable.