}When the IX2-RFACEVA mirror unit cassette is used, the operation in step

1 below is not necessary because no excitation filter is provided with it.

1. Using the push ring driver ², turn the excitation filter push ring @ of the

mirror unit to be mounted in the fluorescence mirror unit counterclockwise
and remove the excitation filter ³.

2. Using the push ring driver ², turn the push ring ƒ of the excitation filter

slider | counterclockwise to remove it.

# The excitation filter should be inserted in the specified orientation.

Insert it so that the indication arrow on the filter frame points the
objective. Ensure the orientation of insertion of the excitation filter
slider into the illuminator before assembly.

3. Drop the excitation filter ³ into the excitation filter slider so that the indica-

tion arrow on the filter frame faces upward (if the up surface comes on the
objective side) and fix with the push ring ƒ.

2

Mounting the Excitation Filter

(Figs. 2 to 4)