Operation, 1 preparation – Olympus IX2-ARCEVA User Manual
Page 15
IX2-ARCEVA
1 2
OPERATION
4-1
Preparation
1
Centering the Burner
(Fig. 5)
}The following settings are the original settings required for this system.
The details may be variable according to the instruction manual for each
lamp housing.
@ Field iris diaphragm: Pulled-out position (Closed)
² Slit:
Pulled-out position (Slit OUT)
³ Wedge prism:
Nearest position (Vacant position)
| Excitation filter:
B or IB excitation filter engaged. Also the B or IB
mirror unit should be engaged in the light path.
Fig. 5
@
2
Adjusting the Slit
(Figs. 6 to 8)
}Open the field iris diaphragm @, set the slit selector lever ² to the OUT
position and set the wedge prism ³ to the vacant position before the
following adjustment. (Fig. 6)
1. Engage the PlanApoN60XO
TIRFM
or PlanApo60XO
TIRFM
objective (which re-
quires immersion oil) in the light path.
2. Place a fluorescent specimen on the stage and bring the specimen into
focus using general fluorescent light of B or IB.
3. Remove the specimen, rotate the revolving nosepiece and screw the
target into an objective position.
Fig. 6
@
³
|
²
²
³