1 3
Fig. 7
4. Push in the slit selector lever ² (IN position) and set the wedge prism
slider ³ to the intermediate click position to engage the slit and wedge
prism in the light path. (Fig. 7)
5. Using the Allen wrench (width across flats 3 mm), turn the two centering
screws | alternately so that a crescent slit is visible on the target.
Adjust so that the slit is located as close as possible to the perimeter of
the target.
6. If the slit is not sharply visible, adjust the focus adjustment screw ƒ using
the Allen wrench (width across flats 3 mm).
²
³
ƒ
|