Simplified Transmitted Polarized Light Observation

}The SZX2-AN analyzer is required additionally.

The SDPlApo1.6XPF/2XPFC objectives cannot be used in the polarized
light observation.

1. Loosen the two analyzer clamping knobs @, fit the analyzer into the

objective tip ² and tighten the clamping knobs (by positioning them on
both sides).

2. Without a specimen, rotate the analyzer rotation ring ³ to make the field

of view darkest (cross-Nikol position).

3. Place a specimen on the rotary stage and rotate the stage to begin

polarized light observation.

@