Post processing, Dynamic data, 3 post processing – RayTek 54301 User Manual
Page 79: Ynamic, Programming guide, 5 dynamic data, Mid 71
Programming Guide
11.4.3 Post Processing
The following parameters can be set to determine the post processing
mode, see section 6.4
P=5
peak hold, hold time: 5 s
F=12.5
valley hold, hold time: 12.5 s
G=10
averaging, average time (90%): 10 s
XY=3
advanced peak hold, hysteresis: 3 K
XY=‐2
advanced valley hold, hysteresis: 2 K
Advanced Peak/Valley Hold with Averaging:
C=250
threshold: 250°C
AA=15
averaging time (90%): 15 s
11.5 Dynamic Data
All temperature related information is calculated 128 times a second.
To request the dynamic data, following commands are available:
?T
target temperature
?I
internal temperature of the sensing head
?XJ
internal temperature of electronics housing
?Q
energy value of the infra‐red temperature
?XT
trigger set point (active/inactive) for the FTC3 input
To check for resets (e.g. power shut down) use the command XI.
Notice, after a reset the unit is new initialized.
?XI
asks for the reset status
!XI0
no reset occurred
!XI1
a reset occurred, new initialization of the unit
XI=0
sets the reset status back to 0
MID
71