Programming Guide 

11.4 Device Setup 

11.4.1 Temperature Calculation 

U=C 

unit for the temperature value 

E=0.950 

Emissivity  setting  (Caution:  according  to  the  settings 
for  “ES”,  see  section  11.4.2 

Emissivity  Setting  and 

Alarm Setpoints

 on page 69.) 

XG=1.000 

Setting for transmission 

For  the  calculation  of  the  temperature  value,  it  is  possible  to  set  an 
offset (relative number to be added to the temperature value), and a 
gain value. 
DG=1.0000 

Gain adjustment for the temperature signal 

DO=0 

Offset adjustment for the temperature signal 

In case the ambient temperature is not requested by the internal head 
temperature,  you  must  set  the  ambient  temperature  values  as 
follows: 
A=250.0 

Ambient temperature (example) 

AC=1 

Control ambient background temp. compensation 

11.4.2 Emissivity Setting and Alarm Set points 

The device allows three choices for the emissivity setting and two for 
the alarm output setting. 

ES 

Selection of the emissivity setting. 

ES=1 

Emissivity  set  by  a  constant  number  according  to  the 
„E“ command 

ES=E 

Emissivity set by a voltage on FTC1 (analog input) 

ES=D 

Emissivity  set  by  the  entries  in  a  table  (selected  by 
digital inputs FTC1 – FTC 3) 

?CE 

asks  for  the  emissivity  value  that  is  actually  used  for 
temperature calculation 