Technical Data
3 Technical Data
3.1 Measurement Specifications
Temperature Range
LT
‐40 to 600°C (‐40 to 1112°F)
for J‐Thermocouple: ‐25 to 600°C (‐13 to 1112°F)
Spectral Response
LT
8 to 14 μm
Response Time
All models
150 ms (95% response)
Accuracy
LT
± 1% or ± 1°C (± 2°F) whichever is greater
LT
± 2°C (± 4°F) for target temp. < – 20°C (4°F)
TC outputs
± 1% or ± 2.5°C (± 5°F) whichever is greater
At ambient temperature 23°C ±5°C (73°F ±9°F)
Repeatability
All models
± 0.5% or ± 0.5°C (± 1°F) whichever is greater
4
MID