beautypg.com

Emissivity setting (digital controlled), 2 emissivity setting (digital controlled), Installation – RayTek 54301 User Manual

Page 33: Mid 25

background image

Installation 

5.4.2 Emissivity Setting (digital controlled) 

The sensor’s electronics contains a table with 8 pre‐installed settings 
for emissivity. To activate these emissivity settings, you need to have 
the  inputs  FTC1,  FTC2,  and  FTC3  connected.  According  to  the 
voltage  level  on  the  FTC  inputs,  one  of  the  table  entries  will  be 
activated. 

0 = Low signal (0 V) 
1 = High signal (5 V) 

A non‐wired input is considered as “High”! 

Table entry Emissivity

(Examples)

FTC3 FTC2 FTC1

0
1
2
3
4
5
6
7

1.100
0.500
0.600
0.700
0.800
0.970
1.000
0.950 

0
0
0
0
1
1
1
1

0
0
1
1
0
0
1
1

0
1
0
1
0
1
0
1

Figure 15: Digital Selection of Emissivity with FTC Inputs 

The  values  in  the  table  can  only  be  changed  by  means  of  the 
DataTemp MultiDrop software. 

MID 

25