Documenting co2 test results, Documenting co2 test results -30, Test results – Philips 862478 User Manual
Page 80

Performance Procedures
4-30
Testing the Monitor
Documenting
CO
2
Test
Results
The following table lists the CO
2
tests and summarizes how to document the test results.
Test
Expected Test Results
What to record on service record
Barometric Pressure Check
X1 = difference between the reference
pressure and the measured ambient
pressure displayed on the monitor
(-12 mmHg < X1 < +12 mmHg)
PCO
2
: P/X1/X2/X3/X4/X5/X6/X7
or
PCO
2
: F/X1/X2/X3/X4/X5/X6/X7
P = passed, F = failed
X1: xx (two digits)
X2: x.x
X3: x.x
X4: xxx
X5: x.x
X6: x.x
X7: x.x
Leakage Check parts 1 and 2
X2 = value of part 1 leakage check on
flowmeter (X2 = 4.0 ml/min)
X3 = value of part 2 leakage check on
flowmeter (X3= 4.0 ml/min)
Pump Check
X4 > 120 mmHg below the ambient
pressure
Flow Rate Check
X5 = CO
2
flow rate (42.5 ml/min <65 ml/min). CO 2 Gas Measurement Calibration Check X6 = difference between measured 2 value and calculated value, based on 5% CO 2 cal. gas (X7<2.6 mmHg, or 0.35 kPa) Calibration Verification X7 = difference between measured 2 value and calculated value, based on 10% CO 2 cal. gas. (X7±{0.07 x value calculated})
CO
CO