Functionality Assurance Tests
Testing the Monitor
4-13
Step
Action
1
Connect SpO
2
simulator cable to the SpO
2
adapter cable. Connect the cable to
the SpO
2
patient monitoring input connector.
2
Set the simulator as follows:
3
a. Power the monitor On. Verify that the values are within the following
tolerances:
– Oxygen Saturation Range 79% to 83%
– Heart Rate Range 67 to 73 bpm
4
Press the Alarm Silence button to temporarily silence the audible alarm.
5
Verify that the heart rate tone source, found in the Heart Rate Menu, is set to
SpO
2
.
6
Press the Volume button on the monitor’s front panel. Within 3 seconds of
having pressed the button, rotate the navigation wheel clockwise to verify that
the beeping heart rate tone sound level increases.
7
Wait 3 seconds. Press the Volume button and rotate the wheel counter-clockwise
and verify that the beeping heart rate tone decreases until it is no longer audible.
Rotate the wheel clockwise to return the beep volume to a comfortable level.
After 3 seconds with no wheel activity, the volume adjust function terminates.
Item
Setting
Oximeter
Nellcor
SpO
2
%
81
BPM
70
Pulse Mod
5.00%