Trends, Trends -10 – Philips 862478 User Manual
Page 182
Measuring and Displaying Parameters
12-10
Specifications
Trends
Parameter
Specification
Type
Graphical and tabular
Memory Storage
12 hours, nonvolatile
Data interval
20 seconds (stored data points is the
average over 20-second interval)
Graphical Trend
Format
One graph per vital sign
Display range
2 hours, scrollable
Vertical scaling
Heart Rate
50-100, 0-250, 25-125 b/min
NBP
50-100, 0-150, 0-300 mmHg
SpO
2
0-100, 60-100, 80-100%
CO
2
0-100, 0-60, 20-50 mmHg
Respiration Rate
0-20, 0-50, 0-150 b/min
Temperature
o
C
15-45, 33-41, 35-39
o
C
Temperature
o
F
55-115, 91-107, 96-102
o
F
Tabular format
One table for all variables.
Six fields per row (time and five vital
signs)
Display interval
Per NBP measurement, or 15 minutes for
no NBP, or 20 seconds during alarm
condition