Testing Safety Information

4-2

Testing the Monitor

Testing Safety Information

Warning

Warning

Perform all leakage tests any time the unit is opened.

AC mains voltage is present on the applied part terminals during this test. Exercise

caution to avoid electrical shock hazard.

Note

The monitor must be placed in Diagnostic Mode, with the NBP test screen active for each of
the NBP tests.

If the test fails, ensure the integrity of the cuff and tubing, then test again. If the test fails
again, verify the integrity of all the pneumatic tubing inside the monitor.

Before beginning the Battery Performance test, ensure that the monitor is disconnected from
the AC power source.