Measuring and displaying parameters, Measuring and displaying parameters -4, Ecg -4 – Philips 862478 User Manual
Page 176
Measuring and Displaying Parameters
12-4
Specifications
Measuring and Displaying Parameters
ECG
Parameter
Specification
Heart Rate Range
20-250 b/min
Heart Rate Accuracy
±5 b/min
Bandwidth: Normal Monitoring
0.5 Hz - 40 Hz
Bandwidth: Extended Low Frequency Range (use
selectable)
0.05 Hz - 40 Hz
Leads
3 lead, user selectable
Display Sweep Speeds
12.5, 25, and 50 mm/sec
Pacemaker Detection
Indicator on waveform display, user
selectable
ECG size (sensitivity)
0.5, 1, 2, 4 mV/cm
Lead-off Detection
detected and displayed
Input Impedance
>5 M
Ω
CMRR (Common Mode Rejection Ratio)
>90 dB at 50 or 60 Hz
Input Dynamic Range
±5 mV AC, ±300 mV DC
Defibrillator Discharge
<5 sec per IEC 60601-2-27
Recovery
<8 sec per AAMI EC13-1992