COMPARING THE InDuo™ METER AND LABORATORY RESULTS

59

• Never use your InDuo™ meter with

blood that has been collected in a
test-tube with a grey-coloured top.

• Use only fresh capillary blood

obtained from the fingertip.

Other reasons for variation
Besides the factors listed in the
OneTouch

®

Ultra Test Strip package insert

there are other reasons for variations:

• Blood glucose levels can change

significantly over short periods,
especially if you have recently eaten,
exercised, taken medication, or
experienced stress

1

.

• If you have eaten recently, the blood

glucose level from a fingertip blood
sample can be up to 3.9 mmol/L
(70 mg/dL) higher than blood drawn
from a vein (venous sample) used for
a laboratory test

2

. It is therefore best

to fast for 8 hours before doing a
comparison test between an InDuo™
meter result and a laboratory result.

• Physical factors, such as the amount

of red blood cells in the blood (a high
or low hematocrit) or the loss of
body fluids (severe dehydration) may
also cause an InDuo™ meter test
result to be different from a
laboratory result.

References
1. Surwitt, R.S., and Feinglos, M.N.:

Diabetes Forecast (1988), April, 49-51

2. Sacks, D.B.: “Carbohydrates”. Burtis,

C.A., and Ashwood, B.R. (ed.), Tietz
Textbook of Clinical Chemistry
.
Philadelphia: W.B. Saunders Company
(1994), 959

DCO_06153701A_ID_OB_CAN 9/6/02 11:36 AM Page 59