Lifescan In Duo User Manual

Page 18

background image

b

c

Apply control solution
• Shake the control solution vial well.
• Remove the cap.
• Squeeze the vial and discard the first drop and

wipe off the dispenser tip with a clean tissue to
obtain an accurate result.

• Squeeze the vial again to get a small hanging drop.
• Apply the hanging drop b to the top edge area of

the OneTouch

®

Ultra Test Strip (where it meets the

narrow channel) until the drop is drawn into the
OneTouch

®

Ultra Test Strip, the confirmation

window is filled, and the meter display begins to
count down.

Read test results
• The meter will count down from N to A and the

control solution test result will appear c.

• Compare the result with the range printed on the

OneTouch

®

Ultra Test Strip vial label. The result

should fall within this range.

18

GETTING YOUR InDuo™ SYSTEM READY FOR USE