Kodak 6B6640 User Manual
Page 9
Calibrating the Printer
March 2001
1-5
2. Click
Yes.
The display shows the Choose Start LUT dialog box.
NOTE: The highlighted LUT in the dialog box above is the most recent
calibration LUT.
3. Click
either
Load Selected LUT or Load Custom LUT (or Cancel to cancel
the calibration process).
If you click Load Selected LUT, the system automatically downloads the
highlighted LUT to the printer to create a test print. When the test print has
been sent, the Scanning Processed Print with Densitometer status icon on
the LED Printer Calibration window highlights.
Continue with
“Obtaining Densities” on page 1-6
4. If you selected Load Custom LUT in step
From the Open dialog box, select or type the name of the LUT file you want
and click Open.
The system automatically downloads the LUT file to the printer.
When the test print has been sent to the printer, the Scanning Processed
Print with Densitometer status icon on the KODAK LED Printer Calibration
window highlights.