Kodak 6B6640 User Manual
Page 30
Troubleshooting
2-14
March 2001
<27>
BAD_PATCH
If the Pattern Recognition comparison window is set to 0 or 1, this error
may result when a single patch on the strip is bad due to dirt, smears,
creases, etc. (See explanation of CW command for more information.)
<30>
BAD_STRING_LENGTH
A strip definition string received as part of the DS (Define Strip) RCI
command did not contain the correct number of characters.
<31>
BAD_CHARACTER
A non-printable character was received in one of the name fields (strip
name or pass name). Printable characters are in the range 0x20 to 0x7E,
inclusive.
<32>
BAD_MEAS_TYPE
The value for the measure type field was greater than 3. Valid values are
0 through 3.
<33>
BAD_COLOR
The value for the color field was greater than 9.
<34>
BAD_STEPS
The value for the number of steps was greater than 72.
<35>
BAD_STOP_LOCATION
The value for the guide setting was less than 4 or greater than 30.
<36>
BAD_OUTPUT_TYPE
The value for the output type was greater than 5.
<37>
MEMORY_ERROR
The definition could not be stored.
<38>
BAD_N_FACTOR
The n-factor (used in Dot calculations) is either less than 0500 (0.500) or
greater than 4500 (4.500).
<39>
STRIP_DOESNT_EXIST
The SD command was issued and the requested strip does not exist.
<3A>
BAD_MIN_MAX_VALUE
The value for the min/max field was greater than 4.
<40>
BAD_SERIAL_NUMBER
An attempt was made to redefine one of the stored serial numbers with a
number containing one or more invalid digits.
Code
Message
Explanation