Densitometer errors, Densitometer errors -12 – Kodak 6B6640 User Manual
Page 28
Troubleshooting
2-12
March 2001
Densitometer Errors
Code
Message
Explanation
<01>
BAD_COMMAND
The densitometer does not recognize a command.
<02>
PRM_RANGE_ERROR
One or more data or address parameters is out of range. See the
command guide for parameter limits.
<03>
DISPLAY_OVERFLOW_
ERROR
An attempt was made to read or write beyond the end of the display. Valid
addresses for the display range from 0x00 to 0x20 (0 to 31 decimal).
<04>
MEMORY_OVERFLOW_
ERROR
An attempt was made to read or write beyond the end of the external
RAM within the densitometer. Valid addresses for the external RAM range
from 0x0000 to 0x7fff (0 to 32767 decimal).
<05>
INVALID_BAUD_RATE
The parameter included with the BR command was invalid. Valid
parameters are 06 (1200), 0C (2400), 18 (4800), 30 (9600), and 60
(19200).
<06>
INVALID_PHOTO_STRIP
The specified photographic control strip does not exist.
<07>
TIMEOUT
The time interval between characters has exceeded 5 seconds.
<08>
PRM_SYNTAX_ERROR.
An invalid character was found in the command string.
<09>
INVALID_PASS
The requested pass does not exist for the currently stored strip (e.g.,
specifying pass 5 for a 4-pass strip).
<0a>
INVALID_STEP
The requested step does not exist for a given pass in the currently stored
strip (e.g., specifying step 25 or step 0 on a 21-step pass).
<0b>
NO_DATA_AVAILABLE
Data was requested but no data is stored in memory.
<0c>
MISSING_PARAMETER
One or more required parameters was omitted.
<10>
LAMP_MARGINAL
The lamp output is at approximately 50% of its original brightness. Strip
readings can still take place, but to maintain proper accuracy and
repeatability, replace the lamp as soon as possible.