Starting the device calibration software, Starting the device calibration software -2 – Kodak 6B6640 User Manual
Page 6
Calibrating the Printer
1-2
March 2001
You need to calibrate the printer when you start the system up each day. You also
need to calibrate the printer when
•
you change paper
•
print quality is questionable
•
the temperature at the site changes more than 5° F (2.8° C)
•
if running more than one shift, at the beginning of each shift
NOTE: Most windows displayed in this section are from the Windows NT version
of the software. The windows for the Macintosh version of the software are
similar.
For information about editing the calibration settings (for example, you may want
to display a calibration graph only if calibration is out of tolerance), see
the Device Configuration” on page 3-25
.
Starting the Device Calibration Software
1.
Open the KODAK PROFESSIONAL Device Calibration application.
The display shows the KODAK PROFESSIONAL Device Calibration window.
If the device you need to calibrate does not appear on the window, you need
to add a calibration device. See
“Adding, Updating, and Deleting Devices” on
.
2. Select the icon for the LED Printer.
3. Click
the
Calibrate icon.
LED Printer
icon