beautypg.com

Kodak 6B6640 User Manual

Page 24

background image

Troubleshooting

2-8

March 2001

910

Configuration file
field:_filename_is invalid.
Calibration will be cancelled.

The user entered an incorrect
value in a configuration file.

• Enter the correct value.
• Reinstall the application.
• If problem continues, call for

service.

911

Error_errcode_occurred.
Processing stopped.

An internal error occurred with the
software.

Restart the application.

919

Error occurred preparing to
handle exit from Kodak
Device Calibration. Will not
perform cleanup at exit.

An internal error occurred with the
software.

Restart the application.

921

Unable to return the printer
to its original state. Fix
printer problem and re-
calibrate.

The application was trying to
cancel but received an error when
sending the original LUT back to
the printer.

Printer is in an unknown state.
Correct any printer problems and
restart the application.

922

There is already a printer
open. Please close the open
printer first.

User tried to open a printer when
the printer was already open.

Close the open printer.

923

An error occurred during the
loading of a printer. Please
check to make sure the file
chosen was of the correct
type and/or format.

An error occurred during the
loading of a printer.

Check that the file chosen was the
correct format or type.

924

No printer loaded.

The operator has not yet opened
a printer.

Open a printer.

941

The height and width values
do not agree with the size of
the target file.

The specified size does not match
the file size.

Enter the correct file size.

942

Unable to configure
densitometer. Densitometer
configuration file may be
corrupt or lost.

The file CalDevDensGen2.cfg is
corrupt or lost.

Restart the application. If problem
continues, call for service.

Code

Status Message

Possible Cause/Subsystem

Possible Solution