Troubleshooting
March 2001
2-9
1003
Unable to load default
configuration file:_filename_.
The default configuration file was
deleted or renamed or moved.
Restore the file manually or
reinstall the software.
1004
Unable to set values
in new configuration file.
An internal error occurred with the
software.
Restart the application.
1005
Unable to save configuration
file:_filename_.
The system is trying to save a file
but there is not enough disk
space.
Remove any unnecessary files to
increase free disk space.
1006
Configuration file field:_field_
is invalid.
The user entered an incorrect
value in a configuration file.
• Enter the correct value.
• Reinstall the application.
1007
Unable to remove
directory:_directory_.
The user does not have the
authority to update the directory.
Verify that the user has update
authority on the directory.
1101
Density DMIN/DMAX points
are invalid.
Density points are invalid.
Reread density values on
densitometer or make a new
calibration print.
1102
Density data is not
monotonic increasing.
Density data is invalid.
Reread density values on
densitometer or make a new
calibration print.
1103
Exceeded maximum number
of allowable data errors.
Density data is invalid.
Reread density values on
densitometer or make a new
calibration print.
1104
Asymmetric Density
Groupings are invalid.
Density data is invalid.
Reread density values on
densitometer or make a new
calibration print.
1201
Unknown response from
densitometer.
During initialization, the
application received an unknown
response from the densitometer.
Verify that the baud rate of the
densitometer matches that of the
applications configuration.
Code
Status Message
Possible Cause/Subsystem
Possible Solution