beautypg.com

Kodak 6B6640 User Manual

Page 16

background image

Calibrating the Printer

1-12

March 2001

If specified in the configuration file, the display shows a list of options.

Re-read the densitometer values is most useful when the graphs show
an unusual plot. Rereading the test print lets you validate the graph, then
returns you to the Out of Tolerance dialog box.

Iterate again using a newly calculated calibration LUT allows you to
keep printing test prints with the newly created LUT without having to save
the LUT table.

Calculate and download a printing LUT and stop: When you select this
option, the LUT is calculated and the file is saved with the creation date
and time as its name.

Cancel and return the printer to its original state saves nothing. Any
corrupted data created after you saved the last time is thrown away.

b. Select one of the four options to try to complete the calibration

successfully.