Hudson Robotics SOLO User Manual
Page 21
21
b. ‘D’ Syringe Type (Side-mounted valved syringes) –
This option allow the user to move liquid from any of the systems fluid
reservoir vessels to or from the tip. For both Aspirate and Dispense, the
reservoir is selected by selecting the appropriate Valve Port in the
displayed input box:
Aspirate – the user can enter a volume (uL) to move from the tip to the
reservoir (even volumes greater than the syringe’s volume), or by
checking the ‘Continuous’ operation box, run the syringe piston
continuously until the user clicks on the “Halt” button. The ‘No.
Strokes’ box can be checked and the desired number of aspiration strokes
entered as another operation option. If there are multiple syringe pumps
on the SOLO, the Syringe box will drop down to allow the user to select
any individual syringe or, at the bottom of the list, “ALL” syringes
operating together.
Dispense – the user can enter a volume (uL) to move from a reservoir to the
tip (even volumes greater than the syringe’s volume), or by
checking the ‘Continuous’ operation box, run the syringe piston
continuously until the user clicks on the “Halt” button. The ‘No.
Strokes’ box can be checked and the desired number of dispensing
strokes entered as another operation option.