5.10.3 Backup Mode
When the FW upgrade is failed or the ITDB-100L is failed to reboot, the ITDB-100L needs
to enter the backup mode and upgrade the FW in the backup mode. Follow the steps
below to enter and exit the backup mode.
Step 1:
Before the ITDB-100L is connected to the power supply, press and hold the
trigger and tuning control button. Refer to Figure 1-3 for the location of the
button.
Step 2:
Connect the ITDB-100L to the power supply. When a beep is generated from the
ITDB-100L, release the trigger and tuning control button. The ITDB-100L is in
backup mode now.
Refer to Section 5.10.2 on how to upgrade the device FW.
Step 3:
To exit the backup mode, remove the power supply, and then reconnect it to the
ITDB-100L.