Calibrating realtime ecg out – Welch Allyn PROPAQ CS User Manual
Page 54
50
Calibration
Welch Allyn Propaq CS Vital Signs Monitor
Calibrating realtime ECG out
1.
Connect the 5-lead ECG cable to the monitor and the simulator.
2.
Set the simulator to 80 bpm, normal sinus rhythm.
3.
Set the monitor to Lead II.
4.
Connect the dc-coupled oscilloscope probe to P9 pin 6 on the Digital PCB, and the
reference probe to TP4 on the Digital PCB.
5.
On the Analog Board, carefully short together U55 pins 1 and 2.
6.
Adjust RP6 on the Analog PCB to 0.0 ± 0.1 V dc.
7.
Remove the short and check for an ECG signal displayed on the oscilloscope of 1.0 ±
0.1 V baseline to peak of the R-wave.
8.
Disconnect the ECG cable from the monitor.
Figure 4. Digital PCB measurement and adjustment locations
digital board layoutdigital board layout
P
9
U
31
P
3
TP
4
Pin 6