beautypg.com

Calibrating realtime ecg out – Welch Allyn PROPAQ CS User Manual

Page 54

background image

50

Calibration

Welch Allyn Propaq CS Vital Signs Monitor

Calibrating realtime ECG out

1.

Connect the 5-lead ECG cable to the monitor and the simulator.

2.

Set the simulator to 80 bpm, normal sinus rhythm.

3.

Set the monitor to Lead II.

4.

Connect the dc-coupled oscilloscope probe to P9 pin 6 on the Digital PCB, and the
reference probe to TP4 on the Digital PCB.

5.

On the Analog Board, carefully short together U55 pins 1 and 2.

6.

Adjust RP6 on the Analog PCB to 0.0 ± 0.1 V dc.

7.

Remove the short and check for an ECG signal displayed on the oscilloscope of 1.0 ±
0.1 V baseline to peak of the R-wave.

8.

Disconnect the ECG cable from the monitor.

Figure 4. Digital PCB measurement and adjustment locations

digital board layoutdigital board layout

P

9

U

31

P

3

TP

4

Pin 6