Fluke patient simulator modification, B - fluke patient simulator modification, Fluke – Welch Allyn PROPAQ CS User Manual
Page 141: Patient simulator modification
B
137
Fluke patient simulator modification
The factory configuration of the Fluke (Dynatech/Nevada) model 214A, 215A, and 217A
Patient Simulators is incompatible with the Propaq CS monitors with IBP channels when
both the ECG and invasive pressure simulator channels are simultaneously connected to
the monitor. Erroneous readings result. The patient simulator must be modified as
described here before you can use both simulator channels with the Propaq CS monitors
with IBP. If only one simulator channel is used, the modification is unnecessary.
Modify the patient simulator's invasive pressure cable as shown in the figure below.
Remove jumpers W1 and W2 from the patient simulator's Top PCB.
Note
Without this modification, the monitor will select a single waveform or parameter
for display.
810-0696-56
3
5
2
4
1