Functional verification
Welch Allyn Propaq CS Vital Signs Monitor
15. Press ECG/RESP > MORE. Then press CHANGE to select the four HR/PR TONE
levels. Verify that the tone volume changes and goes off when turned OFF. Set the
volume to LOW or OFF.
16. Press HOME
to return to the Main Menu.
17. Software versions 3.6x and greater: Press SETUP > MORE > MORE > SERVICE >
YES > MORE > MORE > SETTINGS > ALL ALARMS. Verify alarm status displays
“TURNING OFF ALL AUDIBLE VITAL SIGNS ALARMS”, “ARE YOU SURE?”, YES or
NO.
18. Software versions 3.6x and greater: Press HOME
to return to the Main Menu.
19. Press SETUP > MORE > MORE and check that the displayed time and date are
correct.
Table 5. Keystrokes and monitor responses—Software versions 3.6x and greater
Press
Result
SETUP
Changes the menu.
ALARMS
Changes the display to ALARM STATUS.
4 MIN SUSPND (on menu at bottom of screen)
4 MIN SUSPND changes to RESUME.
Silence/Reset button
(software version 3.6sx and
greater) or Alarm Suspend/Resume (software version
3.5x and earlier).
RESUME changes to 4 MIN SUSPND.
LIMITS
Changes display to ALARM Limits.
NEXT PARAMETER
Moves cursor block down - Next block in parameter limits
NEXT SETTING
Moves cursor block between upper and lower setting.
Home button
Returns to the Main Menu.
NIBP Start/Stop button
twice
Starts/stops the NIBP pump.
Table 6. Keystrokes and monitor responses—All other software versions
Press
Result
SETUP
Changes the menu.
ALARMS
Changes the display to ALARM STATUS.
SUSPEND (on menu at bottom of screen)
SUSPEND changes to RESUME.
Silence/Reset button
(software version 3.6sx and
greater) or Alarm Suspend/Resume (software version
3.5x and earlier).
Toggles Suspend/Resume alarms.
ALL ALARMS (then press NO)
Changes display. Turn ALL OFF. Are You Sure? Y or N.
LIMITS
Changes display to HR Limits.
NEXT PARAMETER
Moves cursor block down - Next block in parameter limits.
NEXT SETTING
Moves cursor block between upper and lower setting.
Home button
Returns to the Main Menu.
NIBP Start/Stop button
twice
Starts/stops the NIBP pump.