Verification and Calibration - B

61

Current Programming and High-Range Measurement Calibration

Action

Display

13.

Connect the appropriate current monitor as shown in figure B-1A.
Connect the DMM (in dc mode) across the current shunt.

14.

Press

Shift Cal

, scroll to CAL CURR, and press

Enter

.

CAL:CURR

15.

Press

Shift Cal

, scroll to CAL LEV, and press

Enter

to select the

first calibration point.

CAL:LEV P1

16.

Press

Shift Cal

and scroll to CAL DATA. Wait for the DMM reading

to stabilize. Then read the DMM and compute the first current value
(DMM reading

÷

shunt resistance). Use the Entry keypad to enter the

first current value.

CAL:DATA 0.00

17.

Press

Shift Cal

, scroll to CAL LEV, use

Ì to scroll to P2 (the

second calibration point), and press

Enter

.

CAL:LEV P2

18.

Press

Shift Cal

and scroll to CAL DATA. Wait for the DMM reading

to stabilize. Then read the DMM and compute the second current
value (DMM reading

÷

shunt resistance). Use the Entry keypad to

enter the second current value.

CAL:DATA 0.00

Negative Current Programming Calibration

Action

Display

19.

Connect a negative supply to the output in series with the appropriate
load resistor as shown in figure B-1C. Set the DMM to operate in
current mode. Set the negative supply to the full scale voltage rating
of the UUT.

20.

Press

Shift Cal

, scroll to CAL CURR NEG, and press

Enter

.

CAL:CURR:NEG

21.

Wait for the dc source to compute the negative current calibration
constant. The display returns to Meter mode when the calculation is
complete.

Low-Range Current Measurement Calibration

Action

Display

22.

Press

Shift Cal

, scroll to CAL CURR MEAS LOW, and press

Enter

.

CAL:CURR:MEAS:LOW

23.

Connect the UUT as shown in figure B-1B with the appropriate load
resistor for your model. Set the DMM to operate in current mode.

24.

Press

Shift Cal

, scroll to CAL LEV, and press

Enter

to select the

first calibration point.

CAL:LEV P1

25.

Press

Shift Cal

and scroll to CAL DATA. Wait for the DMM reading

to stabilize. Then use the Entry keypad and enter the current reading
displayed on the DMM.

CAL:DATA 0.00