Front Panel Operation - 5

47

7 - Saving and Recalling Operating States

NOTE:

This capability is only available when the unit is set to the SCPI programming
language.

You can save up to 4 states (from location 0 to location 3) in non-volatile memory and recall them
from the front panel. All programmable settings are saved.

Save an operating state in location 1 as follows:

Action

Display

1.

Set the instrument to the operating state that you want to save.

2.

Save this state to location 1. Press Save, 1, Enter.

*SAV 1

Recall a saved state as follows:

Action

Display

1.

Recall the state saved in location 1 by pressing Recall, 1, Enter

*RCL 1

Select the power-on state of the dc source as follows:

Action

Display

1.

On the Function keypad, press Output, and scroll through the Output
menu until you get to the PON state command.

PON:STATE RST

2.

Use the ã and ä keys to select either RST or RCL0. RST sets the power-
on state of the unit as defined by the *RST command. RCL0 sets the
power-on state of the unit to the state saved in *RCL location 0.

Clear the non-volatile memory of the dc source as follows:

Action

Display

1.

On the Function keypad, press Output, Enter. This returns the unit to
the factory-default settings.

*RST

2.

Save these settings to location 1. Press Save, 1, Enter.

*SAV 1

3.

Repeat step #2 for memory locations 2 through 4..

*SAV 2
*SAV 3
*SAV 4