To check the power supply voltages – HP 16500B User Manual
Page 67
To check the power supply voltages
Refer to chapter 6, "Replacing Assemblies," for instructions to remove or replace covers and
assemblies.
W A R N I N G
Hazard voltages exist on the power supply, the CRT, and the CRT driver board. This
procedure is to be performed by service-trained personnel aware of the hazards involved,
such as fire and electrical shock.
1
Turn off the instrument, disconnect the power cord, then remove the top cover.
2
Apply power to the instrument.
3
Using a DVM, measure the power supply voltages.
4
Note problems with the power supply, then return to the power supply flowchart.
Power Supply Voltages
Pin
Supply/Signal
Voltage
Pin
Supply/Signal
Voltage
1-5
5 volt
+5 V
21
Ground sense
0 V
6
+5 volt remote sense
+5 V
22
Power valid
+5 V
7-9
+3.5 volt
+3.5 V
23
−
3.25 volt remote sense
−
3.25 V
10
+3.5 volt remote sense
+3.5 V
24-30
−
3.25 volt
−
3.25 V
11
+12 volt
+12 V
31
−
5.2 volt current limit
+2.5 V
12
−
12 volt
−
12 V
32
−
5.2 volt remote sense
−
5.2 V
13
Not connected
33-37
−
5.2 volt
−
5.2 V
14-20
Ground
0 V
38
Remote on/off
+12 V
Troubleshooting
To check the power supply voltages
5–24