To check the power-up self-tests – HP 16500B User Manual
Page 61
To check the power-up self-tests
The logic analysis system automatically performs power-up self-tests when you apply power
to the instrument. The revision number of the operating system shows in the upper-right
corner of the screen during these power-up tests. As each test completes, either "passed" or
"failed" appears on the screen in front of the name of each test.
1
Disconnect all inputs, then insert a formatted disk into the flexible disk drive.
2
Let the instrument warm up for a few minutes, then cycle the power by turning off
and turning on the power switch.
If the instrument is not warmed up, the power-up test screen will complete before you can
view the screen.
3
As the tests complete, check if they pass or fail, then return to the flowcharts.
The Flexible Disk Test reports No Disk if a disk is not in the disk drive. Refer to the tables
below for the screen listing after the tests are complete and for power-up error messages.
Power-Up Self-Test Screen Listing
Performing Power-Up Self-Tests
passed
ROM Test
passed
RAM Test
passed
Interrupt Test
passed
Display Test
passed
HIL Controller Test
passed
Front-Panel Test
passed
Touchscreen Test
passed
Correlator Test
passed
Hard Disk Test
passed
Flexible Disk Test
Loading Module Files
Error Messages at Power-Up
Message
Cause
Remedy
FATAL ERRORS ENCOUNTERED,
BOOT HALTED
A power-up test failed.
Follow the troubleshooting
flowcharts.
SYSTEM DISK NOT FOUND
The mainframe operating system
cannot be found on the hard disk
drive.
Install the operating system
software onto the hard disk drive.
Touchscreen power-up test
impaired
The touchscreen operation is
impaired.
Follow the troubleshooting
flowcharts. The touchscreen might
still operate with a minimal loss of
functionality.
See Also
"The Power-Up Routine" in chapter 8.
Troubleshooting
To check the power-up self-tests
5–18