Exatron 3000b – Extron electronic 3000B User Manual
Page 74
Exatron 3000B
www.exatron.com
4-8
Chapter 4 Handler Setup
A
DRS
D
ESCRIPTION
ROM
RAM
O
PTIONS
C
OMMENTS
00A7
Wiggle Octoload
00
00, FF
To wiggle the octoloader as it moves from track to track,
or during a jam.
00 = Wiggle Octoloader
FF = Wiggle turned off
00A8
Octo 0/4/8 Tubes
08
00, 04,
08
00 setting is for no octoloader. (hand input)
04 sets octoloader for 4-tube input.
08 sets octoloader for 8-tube input.
00A9
Pick-up Delay
08
00 to FF
Time delay to allow the DUT to settle before turning on
the vacuum to pick up the DUT prior to plunging. The
delay counts down in HEX in 10 millisecond steps.
00AA
Stop on Fail Y/N
30
30, FF
See
Display
Diction-
ary
for further
details.
Set to 30: The handler will not stop on fail and the Stop
on Fail set up question is enabled allowing Stop on Fail
to be turned on/off at any time
Set to FF: The handler will stop when a fail sort is
received from the tester. The Stop on Fail set up
question is turned off.
See Address 00BB for Stop on Fail Sort(s) setting
00AB
Gate Open Delay
10
00 to FF
Time for Gate to open on PLCC Type 4 CO Kits
Counts down in HEX in 10 msec steps.
00AC
Package Type
50
28, 40,
11, 12,
50, 5B
28 = DIP CO Kits
40 = PLCC Plunge to Board Type 4
11 = SOIC Type 3 One test site
12 = SOIC Type 3 Two test sites
50 = SOIC Type 4, Plunge to Board
5B = SOIC Type 5B
00AD
D Test ASCII Out
S
S = 53
D = 44
This is the ASCII letter which the E
XATRON
RS-232
interface uses as an output on a double test.
Set to S: Normal. Set to D: Handler will issue S on First
Test Cycle and D on Double Test Cycle.
00AE
Add LCC Option
FF
00, FF
Adds a special delay used by only some PLCC Type 4
CO kits running Thin LCC packages
00 = Turn delay on
FF = Turn delay off
00AF
Add Air Blast
00
00, FF
Adds a “blow off cycle” to the exit cycle on PLCC/SOIC
Type 4 CO Kits only.
00 = turns blow off on
FF = turns blow off off
00B0
P Up Jog Shut RL
00
00 to 19
Jogs shuttle right or left on power up, allows shuttle to
be “aligned” on power up
0? = Jog Left. 1? = Jog Right.
F
OR EXAMPLE
: 02 = Jog Left 2 steps; 13 = Jog Right 3
steps. 9 Steps Maximum.
00B1
P Up Jog
Octo RL
00
00 to 19
Jogs octoloader right or left upon power up, allows
Octoloader to be “aligned” on power up.
0? = Jog Left. 1? = Jog Right.
F
OR EXAMPLE
: 02 = Jog Left 2 steps; 13 = Jog Right 3
steps. 9 Steps Maximum.
00B2
FF