The quantity of recorded data + Smp exceeds the
maximum allowed (35)
Cancel some recordings after transferring
them to a PC
No Unit selected
ERR: SEQ
The Phase Sequence isn’t correct.
Check the Phase Sequence connection.
ERR: P-
The active powers shown on the right side of the message
are negative
If there isn’t a situation of co-generation
check if the clamps are properly connected
ERR: SEQ & P-
The active powers shown on the right side of the message
are negative and the Phase Sequence isn’t correct.
If there isn’t a situation of co-generation
check if the clamps are properly connected /
check the Phase Sequence connection.
ERR: CONNECTION
The instrument has detected a wrong connection to
Voltage inputs
Check the Voltage connections
Error Vref
The user set a Voltage reference not compatible with
voltage at instrument’s input.
Check Voltage Reference set in “CONFIG
RECORDER”
ERR: SYNC
The System Frequency is out of range
Check the System Frequency, check setting
in ANALYZER CONFIG.
Selection Error
There is a mismatch between the Parameter enabled and
the parameter selected for an AUX recording.
Check the parameter enabled in AUX
position and the selected parameter for
recording.
Error1
÷ Error 5
The instrument memory is damaged.
Contact HT Italia assistance