RIGOL M300 User Manual
Page 113

RIGOL
© 2007 RIGOL Technologies, Inc.
User’s Guide for DM3000 Series
3-9
(3). Press Unit ℃, select the unit of the sensor: ℃.
Figure 3- 11 Select Unit
When finish a proper input, press
to save all the changes and back to a higher
level menu.
3. Press Define, the interface shows below:
Figure 3- 12 Define Sensor Data
4. Press Add button, input the first group of reference value: 111.480Ω, 29.5℃.
Figure 3- 13 Input Sensor Data
5. Press SEG Arith Line, Select the arithmetic: Line.
Figure 3- 14 Select Arithmetic
Press
to return (Figure 3-13) and press Done to save all changes, and continue
to input other values.
Figure 3- 15 Add Groups of Data
See also other documents in the category RIGOL Measuring instruments:
- DS1000E (2 pages)
- DS1000E (2 pages)
- DS1000E (2 pages)
- MSO/DS4000 Series (2 pages)
- MSO/DS4000 Series (63 pages)
- MSO/DS4000 Series (45 pages)
- DS6000 Series (79 pages)
- MSO/DS4000 Series (42 pages)
- MSO/DS1000Z Series (2 pages)
- MSO/DS4000 Series (44 pages)
- MSO/DS1000Z Series (12 pages)
- MSO/DS4000 Series (66 pages)
- MSO/DS4000 Series (49 pages)
- DS6000 Series (280 pages)
- DS1000B (9 pages)
- DS1000B (152 pages)
- DS1000E (36 pages)
- DS1000B (166 pages)
- DS1000E (10 pages)
- DS1000E (137 pages)
- DS1000E (166 pages)
- MSO/DS4000 Series (16 pages)
- MSO/DS4000 Series (68 pages)
- MSO/DS2000A Series (2 pages)
- MSO/DS2000A Series (77 pages)
- MSO/DS2000A Series (317 pages)
- MSO/DS1000Z Series (72 pages)
- MSO/DS1000Z Series (8 pages)
- MSO/DS1000Z Series (270 pages)
- M300 (66 pages)
- M300 (169 pages)
- DM3058E (138 pages)
- DM3058E (47 pages)
- M300 (10 pages)
- DM3058E (192 pages)
- M300 (89 pages)
- DSA1030 (12 pages)
- DSA1030 (2 pages)
- DSA1030 (18 pages)
- DSA1030 (2 pages)
- DSA1030 (1 page)
- DSA1030 (2 pages)
- DSA1030 (5 pages)
- DSA1030 (67 pages)